Employee Information  Agency staff, please use your assigned 4 digit ID. Core staff must use UTMB employee ID
Emp ID  First Last
     
Workplace  Using the appropriate selection box below, please select your area of workplace. Note: TDC, Agency, and Medical Students may be selected using the Hospital Service selection box.
 

 UTMB Clinics                  

 
Hospital  Service          
 

 

Click on START TEST to begin

Test Date:     

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1.  Glucose test strips are affected by humidity; the dessicant in lid prevents strips from becoming defective due to exposure to humidity.  It is unacceptable to leave vials uncapped or to leave unused test strips out of their protective, properly capped vial; improperly stored test strips may result in falsely elevated glucose results.

An indication that test strips have been damaged (i.e. by humidity) would be reflected in the control results: Level 1 would be out of range while Level 2 would be within range.

True
False
 


                                                                                            

 

 

 

 

 

2. Use of the Accuchek Inform is limited to screening and monitoring purposes; therefore, a diagnosis of diabetes cannot be made just based on an elevated test result.

True
False

                                                                                            

3. Glucose control solutions are only good for _______ after opening.  The new expiration date for a brand new control set opened on 1/1/07 whose manufacturer expiration date is 3/28/07 is ___________, and should be written on both vials to prevent their use beyond expiration.

6 months; 7/1/06
3 months; 4/1/07
3 months; 3/28/07
9 months; 10/1/07
 

                                                                                            

 

 

 

 

 

 

4. Prior to performing a finger stick, the patient must wash their hands if possible. Then the operator must  wipe the fingertip with alcohol and allow to air dry. Failure to thoroughly clean the fingertip and eliminate any sugary food remnants (for example, after the patient has eaten breakfast) may lead to these substances being measured as part of the patient's blood sugar or a falsely elevated glucose result.

True
False

                                                                                               

5. Maintenance of the Accuchek Inform requires wiping of meter surfaces using a cloth dampened  with _____. Whichever agent is used, one should never spray meter directly, or submerse it in solution.    Also, the meter must never be exposed to high temperatures (i.e. isolettes) for longer than _____.

10% bleach or 70 % isopropyl alcohol BUT NOT Cavicide because it damages the meter's screen. 10 seconds
Cavicide or 10% bleach BUT NOT 70% isopropyl alcohol because it damages the meter's screen.  10 seconds
Cavicide or 70% isopropyl alcohol BUT NOT 10% bleach because it damages the meter's screen.  10 seconds
I give up...way too confusing, we use Cavicide anyways !!!  10 seconds !!

                                                                                               

6. Identify the properly dosed test strip by clicking on Test strips 1, 2, 3, and 4 below.  For improperly dosed test strips, reapplication of sample is acceptable as long as it is done within _______ seconds.

Note: To view test strip image must click on individual blue link below.

test strip 1; 15 seconds 
test strip 2; 15 seconds
test strip 3; 20 seconds
test strip 4; 15 minutes

                                                                                              

7. A critical blood glucose result for an adult is a value that is _____________.  Critical results are documented on the Accuchek Inform using the ______________ Comment Code.

less than 50 mg/dl  or greater than 450 mg/dl; Notified Provider
less than 10 mg/dl  or greater than 600 mg/dl; Notified Provider
less than 30 mg/dl  or greater than 300 mg/dl; Lab Study
less than 40 mg/dl  or greater than 110 mg/dl; Operator error

                                                                                               

8. As of 2005, Joint Commision requires documentation of blood glucose results in the medical record include:

The control lot number in use
The strip lot number in use
The glucometer serial number
The glucose reference (normal) range

                                                                                             

9. Interfering substances are those substances which may lead to lower or higher BG results (depending on substance type).  Manufacturers are obligated to disclose those substances.
As an Accuchek Inform user*, you should be aware of the following interfering substances and how each one impacts the BG result. 

Maltose, Xylose, Galactose  (elevate result)
Estrogens (decrease result)
Bilirubin levels > 20 mg/dl (elevate result)
a & c

* note: Some of the above interfering substances may apply to other glucometer brands .                                                                                         
   

10. Controls must be performed when:

A new lot of strips is opened, or improper storage is suspected (i.e. refrigerated strips, or vial left open)
Once per shift
The meter has been dropped
The patient's clinical symptom's do not really correlate with the BG result
All of the above