CLO test and urease
tablets: 2-8°C
CLO tests are stable for 18 months or until manuf. exp. date. Urease tablets have no exp. at 2-8°C |
MSDS |
|
Cotinine |
|
Cotinine QC Log |
Test strips and control
solutions: 2-8°C
Strips and controls are good for two years or until manuf.
exp. date whichever is first |
|
|
Creatinine, iSTAT |
|
QC Logs |
Analyzer: 10-30°C
Unopened Cartridges: 2-8°C until manuf. exp. date
Cartridges in use may be stored at room temp. for 5 days Controls: 2-8°C until manufact. exp. date or 5 days at room
temp. |
|
|
Glucose, Accuchek Inform |
|
* QC
data captured electronically |
Controls: 15-30°C
Good until manuf. exp. date when unopened. After
opening controls are good for three months. Test Strips: 2-32°C
until manuf. exp. date |
Accuchek Test Strips
Accuchek Control Sol’n
Accuchek Linearity Sol’n |
|
Glucose, Accudata GTS/GTS
Plus |
1-Pager |
* QC
data captured electronically |
See Accuchek Inform
storage conditions |
|
Glucose, Accuchek
Advantage |
|
Advantage QC Log
Cumulative QC Log |
See Accuchek Inform
storage conditions |
|
Glucose, Contour |
|
Elite QC Log
Elite -
Cumulative QC Log
Contour - Cumulative QC Log |
Controls: 15-30°C.
Good until manufacturer's expiration date when unopened.
After opening, controls are good for 6 months or
until the manufacturer's expiration date,
whichever comes first. |
Ascencia Elite Test Strips
Ascencia Elite Controls |
|
Group A Strep (Signify)
SIGNIFY KIT REMOVED
FROM FORMULARY 4/06/11 ALL SITES SHOULD BE USING GENZYME
OSOM KIT |
1-Pager |
QC Log
(note:
use same log to document both internal and onboard controls)
Culture Log
(Rev. 3/24/11) |
|
Strep A Sodium Nitrite |
|
Group A
Strep (Genzyme OSOM) |
|
Internal QC Log
Liquid QC Log
Culture Log (Rev. 3/24/11) |
Entire kit: 15-30°C
Good until manuf. exp. date |
Osom Ultra Strep A |
|
Fecal Occult Blood |
1-Pager |
***Not
Required*** |
Cards and Developer
Solution: 15-30°C
Good until manuf. exp. date |
Guaiac Coloscreen Developer
Kit
Guaiac Hemoccult Sensa
Developer |
|
Rapid Flu (QuickVue A+B)
Rapid Flu (QuickVue A/B) |
|
Patient/Internal QC Log
QuickVue A+B Liquid QC Log
QuickVue A/B Liquid QC Log |
Entire kit: 15-30°C
Good until manuf. exp. date |
QuickVue Flu A+B |
|
Hemoglobin A1C |
Abbreviated Procedure |
DCA 2000 Maintenance Log Rev. 1/04/11
QC Log
Refrigerator Temp Log |
Reagent Cartridge: 2-8°C
or 15-30°C
Reagent cartridges are good until manuf. exp. date when
refrigerated or for 3 months at room temp. Capillary Holder: 2-8 or 15-30°C QC Material Lyophilized: 2-8 or -20°C Reconstituted:
2-8°C
Once reconstituted QC material is good for three months. |
Hgb A1c Controls |
|
Hemoglobin-Hemacue
HB 201+
|
Abbreviated Procedure
1-Pager |
Maintenance Log
QC/Patient Log
Calibration Cuvette
(note: Does NOT apply to HemaCue HB 201 +) |
Cuvettes: 15-30°C
Unopened vials are good until manuf. exp. date. Once
opened, cuvettes are good for 90 days.
Controls: 2-8°C Unopened controls are good until
manuf. exp. date. Once opened, controls are good for 7
days |
Hgb HB201 Cuvettes |
|
Hemoglobin - Hemopoint H2 |
|
Maintenance Log |
Cuvettes: 15-30°C
Unopened cuvettes are good until manuf. exp. date.
Opened cuvettes are good for 90 days.
Controls:2-8°C Open Controls may be
kept at 15-30°C, but are only good for 30 days whereas
opened controls kept at 2-8°C are good for 60 days. |
Hemopoint H2 MSDS |
|
LDX Lipid
Panel |
|
|
Cassettes
must be stored in sealed foil pouch. Cassettes are
good until Manuf. exp.date when refrigerated at 2-8°C.
Cassettes are good for 30 days when stored at room temp.
Control solutions must be stored at 2-10°C. Unopened
vials are good until manuf. exp. date. Opened vials
are good for 30 days. |
|
|
Microhematocrit
HemataSTAT |
1-Pager |
Cell Packing / RPM Chk
|
Controls: 2-8°C
Controls are good for 1 month |
|
|
Microhematocrit- Chronic Home
Dialysis
H & H Sample comparison
procedure |
|
|
|
N/A |
|
Ph-Paper
(new:
pH paper is good for 3 years! see reminding cue in QC log
at right, also policy reflects change) |
1-Pager |
Liquid QC Log
|
pH Paper: 15-30°C
Good for 3 years from open date. Controls: See Urine Dipstick Control |
Urine Dipstick Control |
|
Urine Clinitest |
1-Pager |
Liquid QC Log
|
Tablets: 15-30°C
Good until manuf. exp. date. Controls: See Urine Dipstick Control |
Clinitest Reagent Tablets
Urine Dipstick Control |
|
Prothrombin Time-Coaguchek
XS |
|
|
|
CoaguChek XS |
|
UPT (SiGNIFY)
by
Inverness Medical |
1-Pager |
Internal QC
Liquid QC |
Kit: 15-30°C
Good until manuf. exp. date
Controls: See Urine Dipstick Control |
Urine Dipstick Control |
|
UPT SP Brand Rapid
hCG Cassette |
|
Internal QC Log
Liquid QC Log |
Kit:
2°-30°C (36°-86°F)
Controls: See Urine Dipstick Control |
|
|
UPT Sure Vue |
|
Internal QC
Liquid QC |
Kit: 15-30°C
Good until manuf. exp. date Controls: See Urine Dipstick Control |
Urine Dipstick Control |
|
UPT Genzyme OSOM
Competency sheet for Staff
with existing and current UPT Certification |
|
QC Log
(Combines liquid and Onboard QC documentation) |
Kit: 15-30°C
Good until manuf. exp. date Controls: See Urine Dipstick Control |
Osom hCG Urine Control |
|
UPT Stanbio |
|
Stanbio True QC Log 1
(onboard QC)
Stanbio True QC Log 2
(Liquid QC)
Stanbio QuStick QC Log 1
(onboard QC)
Stanbio QuStick QC Log 2
(Liquid QC) |
Kit: 15-30°C
Good until manuf. exp. date Controls: See Urine Dipstick Control |
StanBio QuStick |
|
Urine
Multistix on Clinitek 50 |
|
|
See Urine Dipstick
Control |
|
|
Urine Dip(UA):
Roche
Chemstrip
Urine Dip (UA):
Siemens/Bayer Multistix |
|
QC Log: Roche
Chemstrip
Refrigerator Temp Log
QC Log:
Siemens/Bayer Multistix
|
Chemstrips: 2-30°C
Multistix: 15-30°C Both types of strips are good until
manuf. exp. date.
Quantimetrix UA
Controls: 2-8°C. If refrigerated controls are
good until expiration date. Controls may also be
stored at 15-30°C, but are only good for 30 days.
These rules apply to opened and unopened bottles! |
Quantimetrix Controls
Bio-Rad qUAntify Controls
UA Chemstrips
Bayer Multistix10 SG |